epBlue Application Runner - Carrying out the application
epBlue™ with MultiCon
English (EN)
31
4.2.7 Entering the filling level of the labware
4.2.7.1 Checking the filling level automatically
If the Detect volumes function is active, the optical sensor checks the filling level after the start.
If the following problems occur, a dialog window appears:
• Filling level too low.
• Filling level too high.
• Filling level cannot be correctly detected.
Abb. 4-7: Checking window
Fig. 4-7: Checking window
If the Calculated volume is larger than the Maximum volume, remove liquid from the vessel.
If the Calculated volume is smaller than the Minimum volume, add liquid to the vessel.
The optical sensor cannot carry out a detection if there is an air bubble in the liquid. Remove the air
bubble. Select Repeat scan to repeat the detection.
Maximum volume
Filling level of the vessel.
Minimum volume
Volume required to remove the set liquid
quantity.
Calculated volume
Volume in the vessel, calculated using vessel
data and the liquid level.
Repeat scan button
Repeat Detect volumes.
Abort button
Aborts the application.
User input button
Manual volume entry.
Accept level and continue button
Accept the displayed calculated volume.
µL input field
Manually select the volume if the User Input
button has been selected.
NOTICE! Dispensing error due to refilling of vessels.
The optical sensor only checks the level of the vessels at the specified location.
Only change the volume of the vessels at the specified location.
Do not change the volume of the vessels at other locations.