epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
168
0x210D Source rack not
defined
Parameter Source was not
edited in the application.
See explanation in
"Cause".
0x210E Source rack not
edited in worktable
The source rack you edited in
the procedure of the
application has been removed
from the worktable.
Edit the rack that you
programmed in the
procedure as Source in the
worktable, or edit a
different source rack in the
application.
0x210F Source rack not
selected in the
labware file window
The source rack you edited in
the application is not selected
or removed in the labware file
window and therefore is not
available for programming.
Select the rack in the
labware file window or edit
a different rack in the
application.
0x2110 Destination rack
not defined
Parameter Source was not
edited in the application.
See explanation in
"Cause".
0x2111 Destination rack
not edited in
worktable
Destination rack was edited in
the procedure of the
application, but it was not
edited in the worktable.
Edit the rack that you
programmed in the
procedure as Destination in
the worktable or edit a
different destination rack
in the application.
0x2112 Destination rack
not selected in the
Labware File
Window
The destination rack you
edited in the application is not
selected or removed in the
labware file window and
therefore is not available for
programming.
Select the rack in the
labware file window or edit
a different rack in the
application.
0x2113 Pattern not defined Parameter Pattern was not
edited in the application.
See explanation in
"Cause".
0x2114 Loading error
(invalid entry in
pattern)
Normally a system error; but
may also be caused by editing
a pattern without destination
positions; or:
File damaged.
Edit a pattern with source
and destination positions.
In other cases:
Call local Eppendorf
Service.
0x211B Liquid type not
defined
Parameter Liquid Type was not
edited in the application.
Choose a "Liquid Type" for
the Liquid Handling
Command.
Code Symptom/message Cause Remedy