epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
172
0x212D Parameter conflict:
8-channel tool
cannot be used for
this source rack
Edit mode/parameter in liquid
handling command:
Source rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24 positions).
Choose another rack or
another tool.
0x212E Parameter conflict:
8-channel tool
cannot be used for
this destination
rack
Edit mode/parameter in liquid
handling command:
Destination rack does not fit
the 8-channel tool (e.g.,
24-well plate or tube rack with
24 positions).
Choose another rack or
another tool.
0x212F Sample number too
large
The number of samples you
have entered will fill more
than one rack (source or
destination, respectively)
based on the programmed
pattern.
Program a pattern that
together with the number
of samples you want to run
will not extend beyond one
rack, or choose a lower
number of samples in the
corresponding Number of
Samples command.
0x2130 Parameter conflict:
tip cannot be used
for this source rack
The Source rack parameter in
the liquid handling command
does not match the selected
tool (e.g., 384-well plate and
TS_1000 or TM1000_8).
Choose another rack or
another tool.
0x2131 Parameter conflict:
tip cannot be used
for this destination
rack
The Destination rack
parameter in the liquid
handling command does not
match the selected tool (e.g.,
384-well plate and TS_1000 or
TM1000_8).
Choose another rack or
another tool.
0x2132 Invalid number of
samples (1 ... 384)
Number of samples you have
entered is too high.
Enter a maximum number
of samples up to 384.
0x2133 Rack or plate not
defined
Edit mode/parameter in
command Transport:
Labware to be transported was
not edited.
See explanation in
"Cause".
0x2134 Rack not selected in
the labware file
window
The rack you edited in a
command Transport
(parameter labware) is not
selected or removed in the
labware file window and
therefore is not available for
programming.
Select the rack in the
labware file window or edit
a different rack in the
application.
Code Symptom/message Cause Remedy