86
A Replacing the O-rings
Hold down the ejection button on the side of the pipette and unscrew the
lower part counterclockwise from the grip.
Using your index finger, press the tool from behind against the nose cone.
Using your thumb, apply pressure from the other side.
The O-ring is severed at one point and can be removed from the nose cone.
Attaching a new O-ring
Attach the assembly aid (shortened pipette tip) onto the nose cone.
Slide the new O-ring over the tip and onto the nose cone.
B Replacing the piston seals
Hold down the ejection key and unscrew the lower part (Fig. 2-6)
counterclockwise from the grip.
With the cone facing downwards, place the pipette lower part horizontally
onto a table and press it down slightly.
Press side B of the tool into the opening (Fig. 2-8) in the housing (Fig. 2-9).
The two stop hooks on the inside are depressed and the cover
plate (Fig. 2-7) pops up!
Push opening A of the tool
(suitable for the pipette size at
hand) from below over the
nose cones in such a way as
to ensure that the sharp edge
in the opening of the tool is
overlapping with the O-ring
(see Fig. 6).
II. Multi-channel model
C
A
B
Fig. 6
Research – Part B – Maintenance
7_ResearchTeil2_E.fm Seite 86 Donnerstag, 17. April 2008 12:34 12