Please see page 3 (Fig. 1) of this manual.
A general description of disassembly and assembly for all single-channel models
is followed by instructions on how to replace specific parts for special pipette
types.
A Disassembly and assembly (Fig. 1 / point 1)
Disassembly
Hold down the ejection button on the side of the pipette and pull off the
ejection sleeve (12) using force.
An exception is the 1 10 mL model:
see Section B "Replacing pistons and seals".
Unscrew the pipette lower part (10) from the grip by turning it to the left
(if necessary, use position A of the wrench provided).
Press the piston holder (1) together and remove it.
The piston is subject to spring tension!
Remove the piston (3) with spring (4) and replace if necessary.
Assembly
Insert the piston (3) with spring (4) into the pipette lower part.
Place the piston holder (1) over the piston, press it together and slide it into the
grooves on the lower part.
Screw the pipette lower part hand-tight into the grip. (Do not use the wrench.)
Hold down the ejection button and re-attach the ejection sleeve.
B Replacing pistons and seals
For fixed- and adjustable-volume (variable) pipettes up to
20 µL
(see Fig. 1 /
point 2):
Disassemble the pipette.
The seal is changed by replacing the entire pipette lower part (10).