10.1 If there is doubt that dispensing data are correct
To avoid dispensing errors, the precision and accuracy of the Reference pipettes
need to be checked regularly. The PICASO II software program (see ordering
information) is available to determine permitted systematic and random
measuring deviation (see Chapter 2 "Technical data"). An SOP (Standard
Operating Procedure) for checking pipettes can be called up from our home page
www.eppendorf.com.
For liquids whose physical data deviate significantly from those of water,
calibration needs to be changed in accordance with Chapter 8.2.3.
Calibration will also need to be changed if the pipette is being used at a very high
altitude.
Eppendorf Reference® – Part A – Chapter 10