Eppendorf Pipette Reference 2 Manual de usuario Pagina 34

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 44
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 33
Technical data
Eppendorf Reference
®
2
English (EN)
34
7.2 Reference 2 Multi-channel variable
Reference 2 single-channel variable
Model Test tip
epT.I.P.S. color code
Volume range
Length
Testing volume Eppendorf AG error limits
Error
Systematic error Random error
± mL ± mL
0.25 mL – 2.5 mL
Increment:
0.002 mL
Red
0.25 mL – 2.5 mL
115 mm
0.25 mL ±4.8 ±0.012 ±1.2 ±0.003
1.25 mL ±0.8 ±0.010 ±0.2 ±0.0025
2.5 mL ±0.6 ±0.015 ±0.2 ±0.005
0.5 mL – 5 mL
Increment:
0.005 mL
Purple
0.1 mL – 5 mL
120 mm
0.5 mL ±2.4 ±0.012 ±0.6 ±0.003
2.5 mL ±1.2 ±0.030 ±0.25 ±0.006
5.0 mL ±0.6 ±0.030 ±0.15 ±0.0075
1 mL – 10 mL
Increment:
0.01 mL
Turquoise
1 mL – 10 mL
165 mm
1.0 mL ±3.0 ±0.030 ±0.6 ±0.006
5.0 mL ±0.8 ±0.040 ±0.2 ±0.010
10.0 mL ±0.6 ±0.060 ±0.15 ±0.015
Reference 2 Variable multi-channel (8-channel/12-channel)
Model Test tip
epT.I.P.S. color code
Volume range
Length
Testing volume Eppendorf AG error limits
Error
Systematic error Random error
± µL ± µL
0.5 µL – 10 µL Medium gray
0.1 µL – 20 µL
40 mm
0.5 µL ±12.0 ±0.06 ±8.0 ±0.04
1 µL ±8.0 ±0.08 ±5.0 ±0.05
5 µL ±4.0 ±0.2 ±2.0 ±0.1
10 µL ±2.0 ±0.2 ±1.0 ±0.1
10 µL – 100 µL Yellow
2 µL 200 µL
53 mm
10 µL ±3.0 ±0.3 ±2.0 ±0.2
50 µL ±1.0 ±0.5 ±0.8 ±0.4
100 µL ±0.8 ±0.8 ±0.3 ±0.3
30
µL – 300 µL
Orange
20 µL – 300 µL
55 mm
30 µL ±3.0 ±0.9 ±1.0 ±0.3
150 µL ±1.0 ±1.5 ±0.5 ±0.75
300 µL ±0.6 ±1.8 ±0.3 ±0.9