4.8 Gateway preferences icon
4.8.1 General tab
The General tab allows the users to name the TCA-3 monitoring system (TCA-3 monitoring systems will be
listed on the screen in alphabetical order), set the sampling rates (how often measurements are made) and
record notes about the TCA-3 monitoring system.
The following sampling rates can be selected:
• Store data every: This option records data if the measurements are within limits which can be slower
• Check for alarm every: This option checks measurements against alarm limits which can be faster
4.8.2 Info tab
The Info tab provides information about the TCA-3 monitoring system and firmware. This information can
be useful when diagnosing system problems remotely.