Operation
TCA-3 Temperature Monitoring System
English (EN)
22
4.6.2 User accounts
The User Accounts tab allows two levels of system access to be created:
Administrator - Administrator can change any parameters such as alarm levels, details of the alarm
contacts, etc.
Reader-only - Reader-only can only view the information from the system, and cannot change any
system parameters.
4.6.3 Default alarm contact
The Default Alarm Contact tab contains a list of E-mail addresses, and/or phone numbers. When enabled,
phone numbers can be called in *sequence rather than all at once by selecting Default List when enabling
an alarm for a sensor.
Alarms will be sent once per event, unless sequential alarming mode is selected.
In *sequential alarming mode, the list will be repeated twice if no user takes responsibility by pressing 5.
If the Administrator forgets their password, contact Eppendorf to have the password reset.